Maintaining the System

5.

Fill the KCl reservoir in the new sensor:

a. Remove the reservoir cap with the hex tool and set it aside as shown in

Figure 3-45.

Figure 3Ć45.āăRemoving the Reservoir Cap with the Hex Tool

Hex Tool

Reservoir Cap

OĆring

OĆring

Sample Path

KCI Reservoir

Vent Hole

b. Partially fill the KCl reservoir as shown in Figure 3-46.

Figure 3Ć46.āăFilling the Reference Sensor

Electrode

Compartment

4M KCl Fill Solution

Fill Line

OĆring

Sample Path

OĆring

c. Tap the front face of the sensor with your knuckle to release any bubbles.

d. Gradually fill the KCl reservoir with KCl fill solution to the fill line.