4Ć69

Troubleshooting the System

5.

If the system does not correctly recognize the sample device, calibrate the
sample door:

a. Press

Calibrate Door

.

b. Insert the door calibration gauge. Use a Bayer Diagnostics 1 mL

syringe size.

c. Press

Start Test

.

d. Remove the sample device when prompted.

e. Check the screen to verify that the sample door is calibrated.

6.

Press

Exit Test

twice.

Temperature/pAtm Test

Use this procedure to test the temperature and barometer controls when the system
generates a D38 or D22 code. If this test is successful, the D38 or D22 code is
cleared from the status area. This test also displays the temperature readings for the
measurement module, the sample path, the measurement module window, the
preheater, and the barometric pressure.

NOTE:

The 37

°

C temperature set point is NIST traceable.

1.

Access the Temperature/pAtm Test from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

2 Temp/pAtm

and press

Enter

.

2.

Press

Start Test

.

3.

Perform the appropriate action.

If you see a . . .

Then . . .

D38 code

a. Check the screen for the message, Temperature control

system on.

b. Press

Stop Test

.

c. If the temperature control system is off, press

Reset

Control

.

d. Continue with step 4.

D22 code

a. Press

Stop Test

.

b. Look at the status area to see if the D22 code has

disappeared.

c. Continue with step 5.

3

Menu Code

2