Maintaining the System

Figure 3Ć65.āăRemoving the Reagent Pump Roller Cage

Platen

Roller Cage

Roller Cage Shaft

Large Tubing

Small Tubing

Tubing Cuff

6.

Clean the roller cage shaft with a lint-free tissue moistened with reagent water
and dry thoroughly.

7.

Place the new roller cage on the shaft.

8.

Turn the roller cage on the shaft until the roller cage stops.

9.

Press the roller cage down until the cage snaps into place.

10. Reinstall the tubing as described in

Replacing the Pump Tubing

, page 3-39.

11. Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

12. Press

Yes

to perform a two-point calibration.