EĆ32

800 Series Operator's Manual

Substance

Concentration Tested

Bilirubin (Direct)

30 mg/dL

Bilirubin (Total)

35 mg/dL

Creatinine

30 mg/dL

Glucose

1000 mg/dL

Hydroxybutyrate

200 mg/dL

Urea

500 mg/dL

Guaiacol

5 mg/dL

Pyruvate

9 mg/dL

Theophylline

9 mg/dL

Penicillamine

25 mg/dL

Isoniazid

2 mg/dL

Uric Acid

10 mg/dL

The following anticoagulants were found not to interfere with lactate recovery at
the indicated concentrations but they cannot be used on the 860 system due to
potential interferences on other sensors or analytes.

Table EĆ27.āăSubstances with Potential Interference with Other Analytes

Substance

Concentration Tested

Citrate

1000 mg/dL

Potassium Oxalate

1000 mg/dL

EDTA

800 mg/dL