Bayer HealthCare Rapidlab 800 User Manual

Page 552

Advertising
background image

EĆ26

800 Series Operator's Manual

Table EĆ21.āă850 Recovery and Precision TestingĊCa

++

Level

Mode

n

WRSD

*

Observed

Expected

Recovery

1

Syringe

21

0.007

0.78

0.76

102.5

Capillary

24

0.011

0.79

0.77

102.1

Microcapillary

24

0.025

0.80

0.75

106.4

Microsyringe

21

0.014

0.72

0.74

97.6

pH/Lytes

18

0.016

0.78

0.76

103.2

2

Syringe

21

0.012

1.87

1.84

102.0

Capillary

24

0.017

1.85

1.82

101.8

Microcapillary

24

0.010

1.65

1.65

99.7

Microsyringe

21

0.008

1.66

1.58

104.8

pH/Lytes

18

0.023

1.64

1.57

104.5

3

Syringe

21

0.028

2.48

2.46

101.2

Capillary

24

0.035

2.56

2.56

100.2

Microcapillary

24

0.077

2.21

2.26

97.6

Microsyringe

21

0.026

2.20

2.14

102.8

pH/Lytes

18

0.042

2.13

2.12

100.6

* WRSD = within-run standard deviation

Advertising