4Ć73

Troubleshooting the System

Sensor Circuit

Expected Value

Lac A

17.602 to 42.398 nA

sample ground/temperature

36.840 to 37.373

°

C

d. Press

Stop Test

.

e. Perform the appropriate action.

If the result . . .

Then . . .

is within the expected
value

continue with step 4.

is not within the expected
value

the sensor circuit fails the test. Continue with step 6.

NOTE:ĄĂ

The measurement test, which requires approximately 8 minutes to

complete, can effectively evaluate measurement system failures.

4.

Install a test/blank sensor:

a. Open the measurement module door.

b. Perform the appropriate action.

If you want to test the . . .

Then install the . . .

pO

2

or pCO

2

sensor circuit

TB1 test/blank sensor.

pH or Na

+

sensor circuit

TB2 test/blank sensor.

K

+

, Ca

++

, or Cl

sensor

circuit

TB3 test/blank sensor.

Glu biosensor circuit

TB4 test/blank sensor.

Lac biosensor circuit

TB4 test/blank sensor.

c. Close the measurement module door.

5.

Perform the Measurement Test:

a. Press

Start Test

.

b. Check the screen for a value of –250 to +250 mV for pCO

2

and –450 to

+450 mV for pH and ISE. If required, remove and reinstall the test/blank
sensor until the value is within this range.

c. For all sensor circuits except pO

2

, wait at least 4 minutes after closing the

door.