Replacing system components, Replacing the reagent bottles – Bayer HealthCare Rapidlab 800 User Manual

Page 237

Advertising
background image

3Ć65

Maintaining the System

Replacing System Components

Replacing the Reagent Bottles

WARNING

Wear safety glasses, gloves, and a laboratory coat when handling the

reagents.

1.

At the Ready screen, press

Enter

.

The Reagent Levels screen appears.

2.

Remove the reagent bottle from the reagent manifold, as shown in Figure 3-37.

Figure 3Ć37.āăRemoving a Reagent Bottle

Reagent

Reagent Fitting

3.

Write the date installed in the space provided on the new bottle.

NOTE:

Do not remove or tighten the cap that contains the reagent septum.

Removing the cap damages the integrity of the reagent septum.

4.

Remove the plug from the cap of the new reagent bottle.

5.

Insert the new reagent bottle into position on the reagent manifold.

6.

Push the bottle to ensure that it fits firmly on the reagent fitting.

7.

Press

Reset Levels

.

The Reset Levels screen appears.

8.

Select the reagent(s) that you replaced and press

Done

.

A prime sequence starts followed by a wash sequence. When the wash
sequence finishes, the Ready screen appears.