3Ć25

Maintaining the System

11. Connect the sample tubing to the sample tubing connector on the measurement

module.

12. Press

Continue

.

A wash sequence starts. When the wash sequence sequence finishes, a
message box appears prompting you to perform a two-point calibration.

13. Press

Yes

to perform a two-point calibration.