Disconnecting the coćox sample path – Bayer HealthCare Rapidlab 800 User Manual

Page 380

Advertising
background image

4Ć92

800 Series Operator's Manual

Disconnecting the COĆox Sample Path

Use this procedure to disconnect the CO-ox sample path from the base model. You
can continue to measure the remaining parameters.

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Turn the CO-ox parameters off:

a. Select

6 System Setup

and press

Enter

.

b. Select

3 Parameters

and press

Enter

.

c. Move to the CO-oximetry frame.

d. Select the tHb parameter and press

Enter

.

This turns all CO-ox parameters off.

e. Press

Next Screen

to access the second screen of parameters.

f. Press

Next Screen

to access the third screen of parameters.

g. Press

Done

when you finish.

2.

Disconnect the CO-ox and base model sample tubing from the sample
connector.

3.

Remove the screw from the sample connector.

4.

Invert the sample connector so that the “Y” is on the bottom as shown in
Figure 4-21.

Figure 4Ć21.ā Disconnecting the COĆox Sample Path

Preheater

Sample Connector

Inverted
Connector

6