Bayer HealthCare Rapidlab 800 User Manual

Page 230

Advertising
background image

3Ć60

Rapidlab 800 Operator's Manual

NOTE:

If your system has a CO-ox module, follow the procedure described for

the appropriate base model. For example, information identified for an 860 also
applies to an 865.

CAUTION:

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

CAUTION:

Do not expose the glucose and lactate biosensors to deproteinizer or

bleach. Replace the biosensors with test/blank sensors (TB4) before cleaning the
sample path. Reinstall the biosensors within 2 hours.

1.

Take the appropriate action.

If you have an . . .

Then . . .

840 or 850

go to step 2.

860

a. Remove the glucose and lactate biosensors.

b. Install the test/blank glucose and lactate sensors

(TB4).

c. Go to step 2.

CAUTION:ĄĂ

Do not expose the reference sensor to bleach. Replace the

reference sensor with the test/blank reference sensor (TB5). Do not substitute a
new reference sensor.

2.

Replace the reference sensor with a test/blank reference sensor:

a. Remove the reference sensor from the measurement module and set it

aside.

b. Install a test/blank reference sensor (TB5) in the measurement module.

3.

Deproteinize the sample path:

a. Prepare the deproteinizer as directed on the package.

b. Select

2 Maintenance

and press

Enter

.

c. Select

1 Deproteinize

and press

Enter

.

d. Invert the deproteinizer vial several times to mix.

e. Insert an aspiration adapter into the sample port and insert the other end

into the deproteinizer.

f. Press

Analyze

.

g. Remove the adapter when prompted.

h. Wait 5 minutes for the deproteinizing cycle to finish.

A message box appears prompting you to condition the sensors.

!

!

860

!

2

Menu Code

1