Defining Significant Digits
Use this procedure to define the number of significant digits reported for certain
primary parameters. Table 5-13 lists the parameters that have high and low
resolution options. The default value is high.
Table 5Ć13.āăReporting Resolution Options
Option
Parameter
Units
Significant Digits
Low
pH
0.01
H
+
nmol/L
1
pCO
2
mmHg
kPa
1
0.1
pO
2
mmHg
kPa
1
0.1
High
pH
0.001
H
+
nmol/L
0.1
pCO
2
mmHg
kPa
0.1
0.01
pO
2
mmHg
kPa
0.1
0.01
1.
Access the System Options screen from the Menu screen:
a. Select
6 System Setup
and press
Enter
.
b. Select
5 System Options
and press
Enter
.
The System Options screen appears with the cursor in the Reporting
Resolution frame, as shown in Figure 5-24.