Maintaining the System

CAUTION:

Do not touch the internal electrode wire. The wire is fragile and is

easily damaged.

3.

Use the hex tool to remove the internal electrode from the cassette you are
replacing.

4.

Discard the cassette according to your laboratory protocol.

5.

Stand the internal electrode on its cap in a safe place.

6.

Remove the new reference sensor cassette from its box.

7.

Fill the electrode compartment in the new cassette:

a. Insert the tip of the KCl fill solution container into the electrode

compartment as shown in Figure 3-48.

b. Slowly fill the electrode compartment by gently squeezing the container until

the KCl fill solution enters the KCl reservoir.

Figure 3Ć48.āăFilling the Electrode Compartment

OĆring

Sample Path

4M KCl Fill Solution

OĆring

Reference Sensor Cassette

(KCI Reservoir)

Electrode

Compartment

CAUTION:

Do not touch the internal electrode wire. The wire is fragile and is

easily damaged.

8.

Use the hex tool to screw the internal electrode into the electrode
compartment, ensuring that you do not cross-thread the electrode.

9.

Fill the KCl reservoir in the new sensor:

a. Remove the reservoir cap with the hex tool as shown in Figure 3-49.

!

!