D35 electronics error, Problem, Solutions – Bayer HealthCare Rapidlab 800 User Manual

Page 323

Advertising
background image

4Ć39

Troubleshooting the System

D35 Electronics Error:

Problem

The system detects an error in the electronics system.

Qualifiers:

1 2

4

5

6

7

8

9

10 11 12 13

Solutions

Qualifier

Procedure

1, 2, 4, 5, 8,
9, 11

perform the Measurement Test as described on page 4-70.

If the measurement test does not remove the D35, shut the system
down and restart it as described in Shutting Down and Restarting the
System
in Section 5.

6 or 7

shut down and restart the system as described in Shutting Down and
Restarting the System
in Section 5.

10

check the temperature control as described in Temperature/pAtm Test,
page 4-69.

12

perform the Valves Test as described on page 4-62.

13

perform the Pump Functions Test as described on page 4-58.

If the D code reappears, contact your Service Representative.

D35

Advertising