Asćrequired procedures, Emptying the waste bottle – Bayer HealthCare Rapidlab 800 User Manual

Page 215

Advertising
background image

3Ć45

Maintaining the System

AsĆRequired Procedures

Emptying the Waste Bottle

Materials required:

S 10% solution of household bleach

NOTE:

Dilute household bleach (5.25% sodium hypochlorite) 1:10 with

reagent-quality water.

S reagent water
S lint-free tissue or swabs

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

WARNING

Wear safety glasses, gloves, and a laboratory coat when performing

this procedure.

1.

Press and hold down the waste bottle latch while removing the waste bottle
from the compartment, as shown in Figure 3-27.

Figure 3Ć27.āăRemoving the Waste Bottle

Waste Bottle

Waste Detector

Waste Outlet Cover

Waste Outlet

Press down the latch.

2.

Cover the waste bottle with the attached cap to avoid spills when transporting
the bottle.