Bayer HealthCare Rapidlab 800 User Manual

Page 549

Advertising
background image

Performance Characteristics

EĆ23

Level

Recovery

Expected

Observed

WRSD

*

n

Mode

4

Syringe

18

2.65

149.6

149.7

99.9

Capillary

22

0.83

150.0

149.7

100.2

Microcapillary

18

0.98

150.0

149.7

100.2

Microsyringe

18

0.62

149.4

149.7

99.8

Expired Gas

18

0.31

149.5

149.7

99.8

5

Syringe

21

1.40

377.4

377.9

99.9

Capillary

21

2.41

377.5

377.9

99.9

Microcapillary

18

5.50

378.9

377.9

100.3

Microsyringe

17

5.51

377.5

377.9

99.9

Expired Gas

17

1.55

379.2

377.9

100.3

* WRSD = within-run standard deviation