Troubleshooting the System

Figure 4Ć15.ā Removing Obstructions from the Sample Port

Capillary Seal

Drip Tray

Sample Door

Retainer Ring

Sample Port

Clot-removal Line

Mount

O-ring

7.

Inspect the mount for obstructions and clean if necessary.

8.

After removing any obstructions, reinstall the sample port:

NOTE:ĄĂ

Ensure that the three O-rings are in place on the mount.

a. Reinstall the sample port onto the mount, matching the tab on the sample

port to the notch in the retainer ring.

b. Grasp the tab on the retainer ring and turn the retainer ring away from you

to lock it in place.

c. Replace the sample door, and ensure that it snaps into place.

9.

Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

10. Press

Yes

to perform a two-point calibration.

11. Analyze QC material as necessary to verify sensor performance.