Patient temperature correction, Gas exchange indices – Bayer HealthCare Rapidlab 800 User Manual

Page 652

Advertising
background image

IĆ32

800 Series Operator's Manual

Patient Temperature Correction

All measurements and calculations are based upon a standard temperature of
37.0

°

C. During sample analysis, you can enter the actual patient temperature value,

which enables the system to provide temperature corrected results. The following
equations, based on NCCLS recommendations,

22

are used:

pH correction

+ DĂpHńDT + ć0.0147 ) 0.0065Ă(7.4 ćpH)

pCO

2

ĂcorrectionĂ +Ă

DĂlogĂpCO

2

DĂT

+ 0.019

pO

2

ĂcorrectionĂ +Ă Ă

DĂlogĂpO

2

DĂT

+

5.49Ă 10

ć11

pO

3.88

2

) 0.071

9.72Ă Ă10

ć9

Ă ĂpO

3.88

2

Ă)Ă2.30

Gas Exchange Indices

Gas exchange indices are a quick way to estimate the relationship between
pulmonary dysfunction and the hypoxia, and to quantitatively determine the degree
of pulmonary shunting. The primary benefit of using gas exchange indices, is that
they are easy to derive at the bedside. However, they do not have a high level of
correlation with the actual measurement of arterial and mixed venous blood and
should be used with discretion. A more reliable method is the Q

s

/Q

t

shunt fraction,

which is based on measurements of pO

2

and oxygen content.

The gas exchange indices are provided with the 800 system for convenience. Final
judgment of their use is in the hands of the physician.

All gas exchange indices require an arterial sample and use measured values at
patient temperature.

Alveolar Oxygen Tension

Alveolar oxygen tension, referred to as pO

2

(A) or p

A

O

2

, is the partial pressure of

oxygen in alveolar gas. It is a primary component in the detection of gas exchange
indices. The following equation

16,28

is used to estimate pO

2

(A).

p

O

2

(A) +

p

I

O

2

Ă *

p

A

CO

2

ĂĂ

ǒ

F

I

O

2

ĂĂ

)

1ĂćĂF

I

O

2

Ă

R

Ǔ

where

p

I

O

2

+

F

I

O

2

Ă Ă (

ptotalĂćĂpH

2

O)

R = gas exchange ratio