Figure 2ć30.ā leveyćjennings chart, Press – Bayer HealthCare Rapidlab 800 User Manual

Page 151

Advertising
background image

2Ć53

Operating the System

Figure 2Ć30.ā LeveyĆJennings Chart

1 2 3 4 5 6 7 8 9 10

11 12 13 14 15 16 17 18 19

20 21 22 23 24 25 26 27 28

QUALITY CONTROL LEVEY-JENNINGS CHART

April 10 1994 16:25

System 850-1001

QC ID 473841 LEVEL 1 Lot 331302

Expiration Date DEC 12 1995

pCO

2

Month APR

88.3

78.3

68.3

58.3

48.3

Y-axis shows
measured values

Accepted Values

Rejected Values

X-axis shows days
of the month

Upper Target Limit
(+2SD)

Lower Target Limit
(-2SD)

Target Value
(mean)

±3SD

NOTE:ĄĂ

The Levey-Jennings chart for the previous month displays only the

data points for QC samples remaining in the file. The Levey-Jennings chart for
the current month displays only the data points for QC samples analyzed to
date.

9.

Perform one of the following tasks.

If you want to . . .

Then . . .

print a Levey-Jennings report

press

Print

. The Levey-Jennings report prints.

view a Levey-Jennings report
for another parameter

press

Previous Screen

. Select a different

parameter and press

Done

.

return to the QC Data Search
Criteria screen or the QC Data
Search Log

press

Done

.

10. Press

Home

to return to the Ready screen.