Printing or transmitting qc sample reports, Figure 2ć28.ā qc sample report, Press – Bayer HealthCare Rapidlab 800 User Manual

Page 147

Advertising
background image

2Ć49

Operating the System

Printing or Transmitting QC Sample Reports

Use this procedure to print or transmit data from the QC Data Search Criteria Log
screen or the QC Data Search Result screen.

1.

Locate the sample you want as described in Recalling QC Data, page 2-45.

2.

Press

Reporting Options

.

If you want to . . .

Then . . .

print a QC sample report

a. Select Print QC Sample Report and

press

Enter

.

b. Press

OK

. Refer to Figure 2-28 for an example of

a QC sample report.

transmit a QC sample
report to an LIS or data
management system

a. Select Send Results and press

Enter

.

b. Press

OK

.

Figure 2Ć28.ā QC Sample Report

. . . . . . .

...........

..................... . . . . . . . . . . . . . . . . . . .

QC SAMPLE APR 12 1994 13:02

SYSTEM 850-1001 QC ID 123

Sequence No 00099 Level 1

Operator ID 12345 Lot 12345

Analysis Date Apr 12 1994 Expiration Dec 31 1995

Analysis Time 11:42

Results Units Target Range

pH 7.613 (7.594 - 7.634)

pCO

2

21.0 mmHg (19.8 - 23.8)

pO

2

148.8 mmHg (142.9 - 162.9)

Na+ 155.9 mmol/L (150.6 - 160.6)

K+ 7.0 mmol/L (6.72 - 7.72)

Ca++ 0.65 mg/dL (0.55 - 0.75)

Cl- 100 mmol/L (95 - 105)

Glucose 41 mg/dL (30 - 50)

Lactate 12.15 mmol/L (9.82 - 13.82)

ctHb 14.4 g/dL (13.5 - 15.5)

FO

2

Hb 1.3 % (0.7 - 1.7)

FCOHb 96.0 % (93.5 - 97.5)

FMetHb 0.5 % (0.2 - 0.6)

FHHb 2.2 % (2.1 - 3.1)

Accepted into QC File 1

QC File Information

File Status

840

850

........................... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

860

These results
appear when an
845, 855, or 865
is interfaced.

..............

.............. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

CO-ox

3.

Press

Home

to return to the Ready screen.

The following result flags can appear on printed reports:

This flag . . .

Indicates . . .

*

the measurement did not reach endpoint.

the result is above the upper limit of the target range.

(Continued)