4Ć67

Troubleshooting the System

4.

Test the fluid detectors with whole blood:

a. Select

Whole Blood

and press

Enter

.

b. Insert a sample device containing a whole blood sample into the sample

port.

c. Press

Start Test

.

d. Remove the sample device when prompted.

e. Check the screen for the message, FD1, FD1A, FD2, FD5 acceptable, to

verify correct fluid detector function.

The system automatically performs a full wash sequence.

5.

Repeat step 4 if you want to test the fluid detectors with another type of fluid.

6.

Press

Exit Test

.

7.

Press

Exit Menu

.

Sample Entry Test

Use this procedure to test the ability of the system to:

S move the sample door and the sample probe
S identify the position of the sample door and the sample probe
S detect the type of sample device
S detect an obstruction
S calibrate the sample door to various types of sample devices
1.

Access the Sample Entry Test from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

1 Fluidics System

and press

Enter

.

c. Select

6 Sample Entry

and press

Enter

.

The Sample Entry Test screen appears as shown in Figure 4-9.

6

1

Menu Code