Bayer HealthCare Rapidlab 800 User Manual

Page 273

Advertising
background image

3Ć101

Maintaining the System

a. Select

2 Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

2.

Ensure that the sample probe is fully retracted.

3.

Remove the old capillary seal:

a. Grasp the top of the capillary seal and pull it toward the right and out of the

sample port, as shown in Figure 3-69.

Figure 3Ć69.āăReplacing the Capillary Seal

Capillary Seal

Sample Port

b. Discard the capillary seal according to your laboratory protocol.

4.

Install the new capillary seal:

a. Hold the seal at the top, angle it toward the right, and push it into the

opening in the sample port.

b. Insert a capillary tube into the port to ensure that the capillary seal fits

securely in place.

5.

Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

6.

Press

Yes

to perform a two-point calibration.