2Ć64

Rapidlab 800 Operator's Manual

Figure 2Ć36.ā TwoĆPoint Calibration Report

2-POINT CALIBBRATION REPORT APR 12 1994 13:02

SYSTEM 860-1001

Sequence no 433

Calibration Date APR 12 1994

Calibration Time 11:42

CAL POINT

New Meas Drift Units

pH 7.382 7.381 -0.001

pCO

2

35.4 34.9 -0.5 mmHg

pO

2

85.0 85.0 0.0 mmHg

Na+ 140.0 140.9 0.9 mmol/L

K+ 4.00 4.00 0.00 mmol/L

Ca++ 1.25 1.26 0.01 mmol/L

Cl- 100 99 -1 mmol/L

Glucose 0 0 0 mg/dL

Lactate 0.00 0.02 0 mmol/L

pAtm 755 mmHg

SLOPE POINT

New Meas Drift Units

pH 6.838 6.839 0.001

pCO

2

70.8 71.3 0.5 mmHg

pO

2

0.0 -0.1 -0.1 mmHg

Na+ 100.0 100.3 0.3 mmol/L

K+ 8.00 8.04 0.04 mmol/L

Ca++ 2.50 2.52 0.02 mmol/L

Cl- 70 70 0.00 mmol/L

Glucose 180 180 0 mg/dL

Lactate 0.00 0.02 0 mmol/L

Atmospheric pressure
appears on all systems

. . . . . . .

...........

....................... . . . . . . . . . . . . . . . . . . .

840

850

............................. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

860

. . . . . . .

...........

........................... . . . . . . . . . . . . . . . . . . .

840

850

............................... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

860

7.

Press

Done

when you finish.

If . . .

Then . . .

more than one report is
found

the Done Options message box appears:

S

Select Next Record and press

OK

to view the next

report that appears on the log.

S

Select Previous Record and press

OK

to view the

previous report that appears on the log.

S

Select Search Criteria Screen and press

OK

to view the

Calibration Data Search Criteria screen.

S

Press

Cancel

to close this message box. The Search

Results screen appears.

one sample is found

the Calibration Data Search Criteria screen appears.

If the selected sample has a D code or status message associated with it, the View
Diagnostics F-key appears. Press View Diagnostics to view the Diagnostics
message box.

Procedural

Notes