Editing qc file setup – Bayer HealthCare Rapidlab 800 User Manual

Page 459

Advertising
background image

System Administration

5Ć49

Editing QC File Setup

Use this procedure to edit QC setup information in an existing QC file under one
of the following circumstances:

S When you change QC lots, you want to replace the old QC file information with

the new lot information.

S You want to change the high and low range values for a parameter.

If you are creating a QC file for the first time, refer to Creating New QC Files,
page 5-46.

NOTE:

During this procedure, if you are changing QC lots, you must delete the

existing QC data stored in the file. If you want to save a copy of the data, archive
the file to a diskette before performing this procedure. Refer to Archiving QC
Data
, page 5-68, for more information.

When you change QC lots, replace the information shown in Table 5-18 with the
data from the new QC material.

Table 5Ć18.āăQC File Information

Field

Description

QC ID

number that identifies the QC material

level

number that describes the level of QC material

lot number

number that identifies the lot

expiration date

date after which you cannot use the QC material

1.

Access the QC File screen from the Menu screen:

a. Select

5 Operating Setup

and press

Enter

.

b. Select

1 QC Setup

and press

Enter

.

c. Select

1 QC Files

and press

Enter

.

2.

Select the file number that you want to edit and press

Enter

.

NOTE:ĄĂ

You cannot select File 13 as a QC file.

3.

Press

Done

.

The QC File Setup screen appears as shown in Figure 5-29.

1

1

Menu Code

5

Advertising