800 Series Operator's Manual

5Ć62

Configuring for External Devices

Use the following procedures to configure the 800 system for any of the following
external devices:

S 270 CO-oximeter
S 800 series compatible ticket printer
S Bayer Diagnostics data management systems
S bar code scanner
S line printer
S laboratory or hospital information system

Table 5-25 lists the ports you can use with each device.

Table 5Ć25.āăPorts and Devices

Select . . .

To connect . . .

Serial port 1

the ticket printer.

Serial port 2 and 3

the 270 CO-oximeter and an LIS or data management system.

Parallel port

other printer types, such as a line printer.

Barcode

the bar code scanner.

1.

Access the Port Selection screen from the Menu screen:

a. Select

6 System Setup

and press

Enter

.

b. Select

6 Communications

and press

Enter

.

The Port Selection screen appears, as shown in Figure 5-38.

Figure 5Ć38.ā Port Selection Screen for an 850

Select the port you

want to use.