Reporting results with calibration drift – Bayer HealthCare Rapidlab 800 User Manual

Page 450

Advertising
background image

800 Series Operator's Manual

5Ć40

Figure 5Ć24.ā System Options Screen for an 850

Move to the Reporting

Resolution frame and

select High or Low.

2.

Select

High

or

Low

and press

Enter

.

3.

Press

Done

.

4.

Define another setup function or press

Exit Menu

to return to the Ready

screen.

The option that you select for the Reporting Resolution is saved with each patient
and QC sample. If you recall the sample data later and have changed the
resolution, the system prints the report using the resolution selected when the
sample was saved.

Reporting Results with Calibration Drift

Use this procedure to define the way the system processes patient sample results
when a sensor is out of calibration. When the system detects a sensor that has
excessive calibration drift, it cannot analyze samples until the drift is corrected. By
selecting Yes for Report Results with Cal Drift, you have the option to let the
system analyze a sample when a sensor is out of calibration.

Table 5Ć14.āăReport Results with Cal Drift Options

OptionDescription

Yes

The system analyzes the sample and logs the event in the status log. The
printed report contains the message, not in calibration.

No

A D2 code appears and the affected sensor is turned off. Analysis continues
but no results are generated for the affected sensor. The default value is No.

Procedural

Notes