Bayer HealthCare Rapidlab 800 User Manual

Page 374

Advertising
background image

4Ć86

800 Series Operator's Manual

Figure 4Ć19.ā Removing Obstructions from the SpringĆLoaded Latch on an 850

Grasp the tab on the
reference sensor and
pull it up and out of the
measurement module.

Clot-removal
Line

Push the spring-loaded
latch to the right.

d. Connect the measurement module tubing to the latch.

e. Reinstall the reference sensor, ensuring that O-rings are in place.

8.

Press the release tab to close the spring-loaded latch.

9.

Ensure that the sensors are aligned correctly.

10. Close the measurement module door.

11. Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

12. Allow the system to warm up at least 15 minutes.

13. Press

Yes

to perform a two-point calibration.

14. Analyze a minimum of two levels of QC material to verify sensor

performance.