Performance Characteristics

EĆ15

Level

Recovery

Expected

Observed

WRSD

*

n

Mode

Capillary

18

0.68

71.7

71.3

100.6

Microcapillary

24

2.24

71.0

71.3

99.5

Microsyringe

24

0.33

71.4

71.3

100.1

Expired Gas

24

0.17

71.1

71.3

99.7

* WRSD = within-run standard deviation

Table EĆ13.āă840 Recovery and Precision TestingĊpO

2

Level

Mode

n

WRSD

*

Observed

Expected

Recovery

1

Syringe

24

0.21

28.2

28.5

99.1

Capillary

18

0.40

28.3

28.5

99.3

Microcapillary

24

1.15

29.0

28.5

101.7

Microsyringe

24

0.17

28.6

28.5

100.3

Expired Gas

24

0.15

28.2

28.5

98.8

2

Syringe

24

0.19

50.0

49.9

100.3

Capillary

18

0.23

50.5

49.9

101.2

Microcapillary

24

0.71

49.0

49.9

98.3

Microsyringe

24

0.15

49.7

49.9

99.6

Expired Gas

24

0.14

49.7

49.9

99.7

3

Syringe

24

0.66

86.6

85.6

101.2

Capillary

18

0.36

84.8

85.6

99.1

Microcapillary

23

1.11

86.2

85.6

100.7

Microsyringe

24

1.08

85.8

85.6

100.3

Expired Gas

24

0.15

85.5

85.6

99.9

(Continued)