Performance Characteristics

EĆ35

Parameter

TotSD

[

WRSD

*

Mean

n

Level

2

177

0.99

0.011

0.049

3

170

0.52

0.008

0.033

* WRSD = within-run standard deviation

[

TotSD = total standard deviation

]

Performance determined with metabolite recal on. Refer to Glucose Biosensor Calibration, in Section 1
for information on performance with recal off.

Table E-30 summarizes the results of the 860 system precision for CVM levels 1
and 4.

Table EĆ30.āă860 CVM Precision Results

Parameter

Level

n

Mean

WRSD

*

TotSD

[

pH

1

565

6.782

0.002

0.004

4

551

7.830

0.007

0.009

pCO

2

1

575

107.6

1.48

3.11

4

548

13.0

0.37

0.74

pO

2

1

563

26.0

1.63

2.02

4

539

249.0

6.05

8.36

Na

+

1

592

89.6

0.65

1.25

4

560

166.1

0.75

1.32

K

+

1

592

1.44

0.036

0.052

4

563

15.34

0.177

0.315

Cl

1

586

134.2

0.95

2.48

4

560

73.0

0.85

1.17

Ca

++

1

592

2.75

0.038

0.065

4

560

0.51

0.012

0.017

* WRSD = within-run standard deviation

[

TotSD = total standard deviation