Bayer HealthCare Rapidlab 800 User Manual

Page 260

Advertising
background image

3Ć88

Rapidlab 800 Operator's Manual

Figure 3Ć61.āăInstalling the Glucose and Lactate Biosensors

Align the contacts on the sensor with the

contacts in the measurement module.

a. Align the contacts on the biosensor with the contacts in the measurement

module.

b. Snap the body of the biosensor down into place. The contacts must be flush

with the biosensor.

c. Press the tab on the spring-loaded latch down to release the latch.

d. Verify that the biosensors are installed from left to right in the following

order in the measurement module:

pO

2

pCO

2

GRD

pH

Na

+

K

+

Ca

++

Cl

Ref

Glu

Lac

6.

Close the measurement module door.

7.

Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

8.

Press

Yes

to perform a two-point calibration.

9.

Perform two additional two-point calibrations to start hydrating the biosensor
as soon as possible:

a. Move to the Main Menu.

b. Select

1 Calibration

and press

Enter

.

c. Select

2 Two-point

and press

Enter

.

10. Allow the biosensors to warm up for at least 30 minutes.

11. Verify that the system temperature is within the acceptable range.

1

Menu Code

2

(from the Main Menu)