Analyzing syringe samples – Bayer HealthCare Rapidlab 800 User Manual

Page 104

Advertising
background image

2Ć6

Rapidlab 800 Operator's Manual

Analyzing Syringe Samples

The following tables list the minimum volumes required by each system to analyze
syringe samples:

Table 2Ć1.āăMinimum Sample Volumes

System

Minimum Sample Volume

840
844, 845

0

90

µ

L

140

µ

L

850
854, 855

110

µ

L

160

µ

L

860
864, 865

125

µ

L

175

µ

L

The system automatically aspirates the sample. If you want the system to analyze a
different set of parameters for the sample, press Change Parameters to select
another panel. Refer to Selecting Parameter Panels, page 2-29, for more
information.

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Prepare the sample and insert the syringe into the sample port, as shown in
Figure 2-2.

Figure 2Ć2.ā Inserting a Syringe Sample

Sample Port

Syringe Sample

CAUTION:ĄĂ

To ensure the accuracy of the CO-ox measurement, close the

CO-ox cover before pressing Analyze.

!