EĆ18

800 Series Operator's Manual

Parameter

TotSD

[

WRSD

*

Mean

n

Level

2

71

133.0

0.21

0.86

3

72

154.5

0.36

1.58

K

+

1

72

2.42

0.007

0.023

2

71

4.96

0.017

0.036

3

72

7.71

0.022

0.087

Cl

1

72

79.2

0.35

0.57

2

71

101.7

0.27

0.53

3

72

121.6

0.29

0.92

Ca

++

1

72

1.54

0.006

0.012

2

71

1.12

0.005

0.011

3

72

0.62

0.004

0.009

* WRSD = within-run standard deviation

[

TotSD = total standard deviation

Precision on aqueous calibration verification materials was estimated using four
850 systems. As many as seven runs per instrument were made over seven days.
Two replicates of each control level were analyzed in each run.

Table E-15 summarizes the results of the 850 system precision for CVM levels 1
and 4.

Table EĆ15.āă850 CVM Precision Results

Parameter

Level

n

Mean

WRSD

*

TotSD

[

pH

1

44

6.806

0.001

0.004

4

44

7.805

0.002

0.004

pCO

2

1

44

103.5

0.93

1.49

4

44

13.4

0.13

0.28

(Continued)