Removing obstructions from the measurement module – Bayer HealthCare Rapidlab 800 User Manual

Page 370

Advertising
background image

4Ć82

800 Series Operator's Manual

Removing Obstructions from the Measurement Module

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Stop the system from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

2.

Open the measurement module door:

a. Push up the door latches located on the lower corners of the module door

and lift up the door.

b. Push the spring-loaded latch to the right.

CAUTION:ĄĂ

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

3.

Inspect the pO

2

and pCO

2

(gas) sensors for obstructions:

CAUTION:ĄĂ

To avoid damage to the sensors, do not attempt to push the

clot-removal line through the gas sensors.

a. Remove the sensors and inspect the sample path for obstructions.

b. Place a piece of tubing, or use a capillary adapter, on the tip of a syringe.

CAUTION:ĄĂ

Do not exert excessive pressure on the gas sensor membrane.

c. Place the syringe against the sensor sample path and inject air through the

sensor as shown in Figure 4-16.

Figure 4Ć16.ā Using a Syringe to Clear a Sensor

Capillary Adapter

Syringe

Sensor

O-ring

2

Menu Code

7

!

!

!

Advertising