2Ć25

Operating the System

1.

Analyze the arterial, and the venous or mixed venous samples you want to use
for a-v studies.

2.

Select the first patient sample for a-v studies:

a. Select

Menu

.

b. Select

4 Data Recall

and press

Enter

.

c. Select

1 Patient Data

and press

Enter

.

The Patient Data Search Criteria screen appears.

3.

Enter the patient ID for the sample and press

Done

.

The Patient Data Search Log screen appears.

4.

Select the patient sample for a-v studies and press

Reporting Options.

The Reporting Options message box appears.

5.

Select

Select for aĆv Studies

and press

OK

.

The First a-v Sample Selected message box appears.

6.

Press

OK

.

If . . .

Then . . .

only one sample is found

the Patient Data Search Result screen
appears.

Press

Reporting Options

. The Reporting

Options message box appears.

1

more than one sample is found

the Patient Data Search Log screen appears.

a. Select the next sample for a-v studies

and press

Enter

.

b. Press

Reporting Options

. The

Reporting Options message box appears.

7.

Select

Select for aĆv Studies

and press

OK

.

The Confirm Samples for a-v Studies message box appears.

NOTE:ĄĂ

Record and use the sequence numbers to access the sample results

from the Patient Data Search Log.

8.

Enter the Qt value if you want VO

2

and DO

2

results.