Glucose and lactate biosensors – Bayer HealthCare Rapidlab 800 User Manual

Page 640

Advertising
background image

IĆ20

800 Series Operator's Manual

Glucose and Lactate Biosensors

The glucose and lactate biosensors are complete electrochemical cells that
incorporate amperometric technology to measure glucose or lactate concentration
in samples. The biosensors consist of four electrodes.

*

The measuring electrode

contains platinum and glucose oxidase or lactate oxidase in a binder, while the
reference electrode is composed of Ag/AgCl. Two other electrodes are also
present. The counter electrode is a Pt (platinum) conductor that ensures a constant
applied potential. Another measuring electrode, without the enzyme, determines
interfering substances in the sample. The potential from interfering substances is
removed from the total differential measurement. A microporous cover membrane
separates the electrodes from the sample.

Figure IĆ10.āăBiosensor

Sample Path

Counter Electrode

Glucose Measuring
Electrode

Reference Electrode

Interference Measuring
Electrode

A constant voltage, called a polarizing voltage, is maintained during analysis. In
the glucose sensor, glucose from the sample interacts with the glucose oxidase on
the surface of the measuring electrode to form hydrogen peroxide and gluconic
acid

C

6

H

12

O

6

) H

2

O

) O

2

GOX

³ C

6

H

12

O

7

) H

2

O

2

(glucose)

(gluconic acid)

where GOX is the glucose oxidase.

The polarizing voltage is sufficient to cause oxidation of the hydrogen peroxide to
oxygen.

H

2

O

2

³ 2H

)

) O

2

) 2e

* Platinized activated carbon electrode technology license from Cambridge Life Sciences plc. under U.S.

Patent Nos. 4,970,145 and 5,160,418 and foreign counterparts.