D21 processing error – Bayer HealthCare Rapidlab 800 User Manual

Page 317

Advertising
background image

4Ć33

Troubleshooting the System

D21 Processing Error:

Problem

An internal communication problem between the system processors has occurred.

Qualifiers:

1

3

Solutions

Qualifier

Procedure

1

Shut the system down and restart as described in Shutting Down and
Restarting the System
in Section 5.

3

Disconnect the CO-ox sample path as described in Disconnecting the
CO-ox Sample Path
, page 4-92. You can continue to analyze samples
on the base model.

Check that the communications cable connecting the CO-ox module
to the base model is installed correctly.

Shut the system down and restart as described in Shutting Down and
Restarting the System
in Section 5.

If the D code reappears, contact your Service Representative.

D21