Bayer HealthCare Rapidlab 800 User Manual

Page 473

Advertising
background image

System Administration

5Ć63

2.

Select the appropriate serial port, press

Enter

,

and then press

Done

.

Refer to Table 5-25 to determine which port to select for your device.

The Device Selection screen appears, as shown in Figure 5-39. If the port
already has a device assigned, that device is selected. If required, press

Previous Screen

and select another port.

NOTE:ĄĂ

You can only select a device that has the diamond symbol before it. If

a device cannot be connected to the selected port the diamond symbol is
absent.

Figure 5Ć39.ā Device Selection Screen for an 850

Select the device you

want to connect to the

port.

3.

Select the required device, and then press

Enter.

If you . . .

Then . . .

select Bar Code

press

Next Screen

. The Bar Code Options screen appears. Select

the appropriate symbology. Select

No Additional Symbology

if

Code 128 is the only symbology you require. Then press

Enter

.

Symbology 128 is always enabled when you install the bar code
scanner. You cannot select or deselect it from the menu. You may
select one more symbology in addition to code 128. Refer to the
800 Series Bar Coding Features technical bulletin for detailed
information on barcode setup. Press

Done

when you finish

.

select LIS/HIS

press

Next Screen

.

The Device Selection screen appears

containing the communication parameters, as shown in
Figure 5-40. Continue to Step 4.

select any other
device

press

Done

.

You are prompted to configure another port. If you

press

No

, the Menu screen appears. If you press

Yes

, the Port

Selection screen appears.