Bayer HealthCare Rapidlab 800 User Manual

Page 231

Advertising
background image

3Ć61

Maintaining the System

Wear safety glasses, gloves, and a laboratory coat when handling bleach.

4.

Perform the cleaning cycle:

a. Press

Yes

.

b. Insert an aspiration adapter into the sample port and immerse the other end

in the 10% bleach solution.

c. Press

Analyze

.

d. When prompted, remove the adapter.

e. Wait 5 minutes for the cleaning cycle to finish.

NOTE:ĄĂ

Press

Cancel

if you want to stop cleaning.

When the cycle finishes, the system performs an extended wash sequence. The
Calibrate System message box appears at the end of the wash sequence.

5.

Press

No

.

6.

Initiate a wash sequence from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

8 Wash

and press

Enter

.

CAUTION:

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

NOTE:

As you install the sensors, ensure that the sensors are in the correct

location. Visually verify that you align the contacts on the sensors with the contacts
in the measurement module.

7.

When the wash sequence finishes, remove the test/blank glucose and lactate
sensors (TB4) and reinstall the glucose and lactate biosensors as described in

Replacing the Glucose and Lactate Biosensors

, page 3-86.

8.

Remove the test/blank reference sensor (TB5) and reinstall the reference
sensor as described in

Replacing the Reference Sensor

, page 3-71.

9.

Empty and clean the waste bottle, waste outlets, and waste outlet cover as
described in

Emptying the Waste Bottle

, page 3-45.

10. Perform a two-point calibration:

a. Select

1 Calibration

and press

Enter

.

b. Select

2 Two-point

and press

Enter

.

11. Analyze a minimum of two levels of quality control material to verify sensor

performance.

WARNING

2

Menu Code

8

(from the Main Menu)

!

860

860

1

Menu Code

2

(from the Main Menu)