System Administration

5Ć61

c. Select

2

Cal Intervals

and press

Enter

.

The Cal Intervals screen appears with the cursor in the Calibration Frequency
frame, as shown in Figure 5-37.

Figure 5Ć37.ā Cal Intervals Screen for an 850

Both interval fields

appear when you select

Fixed Time. Only the

OneĆpoint Interval field

appears when you

select Flexible Time.

Select On if you want

calibrations to repeat

automatically.

2.

Select the options you want.

If you want to change . . .

Then . . .

Calibration Frequency

Select Fixed Time or Flexible Time and press

Enter

.

NOTE:

If you select Flexible Time the interval and

Auto Repeat options are not available.

One-point, Two-point, or
tHb Slope Interval

a. Move to the appropriate field.

b. Type the amount of time you want between

calibrations and press

Enter

.

Auto Repeat

a. Move to the Auto Repeat frame.

b. Select

On

and press

Enter

to automatically repeat

calibrations.
Select

Off

and press

Enter

to stop repeat automatic

calibrations.

Metabolite Recal

a. Move to the Metabolite Recal frame.

b. Select

On

and press

Enter

to perform one-point

metabolite calibrations automatically.
Select

Off

and press

Enter

to stop automatic

one-point metabolite calibrations.

3.

Press

Done

when you finish.

4.

You can define another setup function or press

Exit Menu

to return to the

Ready screen.