Removing and checking the sensors – Bayer HealthCare Rapidlab 800 User Manual

Page 351

Advertising
background image

4Ć65

Troubleshooting the System

d. Press

Stop Test

.

Ensure that the bubbles stop flowing from the adapter.

4.

Check the flow of Slope Gas:

a. Select

Slope Gas

and press

Enter

.

b. If required, push down the sample door halfway open, insert an aspiration

adapter into the sample port, and immerse the other end in the water.

c. Press

Start Test

.

Ensure that a steady stream of bubbles flows from the aspiration adapter.

d. Press

Stop Test

.

Ensure that the bubbles stop flowing from the adapter.

5.

Remove the aspiration adapter and manually close the sample door.

6.

Press

Exit Test

.

7.

Press

Exit Menu

.

When you use the left arrow key to exit the Troubleshooting menu and enter the
Main menu, the system performs a wash.

Removing and Checking the Sensors

1.

Stop the system from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

CAUTION:

Touch the inner surface of the module frame to discharge static

buildup before removing or returning sensors.

2.

Remove the sensors.

3.

Check that the O-rings are in place.

4.

Install the sensors in their correct positions by aligning the top of the sensor
with the sensor contact and snapping the body of the sensor into place.

5.

Close the measurement module door.

6.

Press

Continue

and allow the system to warm up.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

7.

Press

Yes

to perform a two-point calibration.

Procedural

Notes

2

Menu Code

7

!

Advertising