Replacing the glucose and lactate biosensors – Bayer HealthCare Rapidlab 800 User Manual

Page 258

Advertising
background image

3Ć86

Rapidlab 800 Operator's Manual

g. Allow the system to warm up.

h. Press

Exit Test

.

11. When the temperature is stable, perform a two-point calibration:

a. Select

1 Calibration

and press

Enter

.

b. Select

2 Two-point

and press

Enter

.

The Ready screen appears when the calibration finishes.

12. Analyze a minimum of two levels of quality control materials to verify sensor

performance.

After the sensor temperature equilibrates, remove the sensor and inspect for
bubbles. As the temperature of the sensor rises to 37

°

C, gas is driven from the

solution, causing bubbles. Remove any bubbles present.

Replacing the Glucose and Lactate Biosensors

Materials required:

S glucose biosensor
S lactate biosensor

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

NOTE:

Equilibrate the glucose and lactate biosensors at room temperature (18 to

25

°

C) for at least 1 hour before use. Keep the biosensors in their foil packages

while they are equilibrating.

1.

Stop the system from the Menu screen:

a. Select

2

Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

1

Menu Code

2

(from the Main Menu)

Procedural

Notes

2

Menu Code

7