Performing calibrations from the menu mode – Bayer HealthCare Rapidlab 800 User Manual

Page 154

Advertising
background image

2Ć56

Rapidlab 800 Operator's Manual

Performing Calibrations from the Menu Mode

You can perform calibrations from either the Analyze mode or the Menu mode. All
calibrations available on the 840, 850, and 860 are also available on systems with
the CO-ox module—the 844, 845, 854, 855, 864, and 865.

1.

From the Menu screen, select

1 Calibration

and press

Enter

.

The Calibration menu appears, as shown in Figure 2-32.

Figure 2Ć32.ā Calibration Menu for an 850

2.

Select a calibration type and press

Enter

.

NOTE:ĄĂ

The tHb Slope Calibration requires you to enter a value and to

introduce the slope reagent. Refer to Performing tHb Slope Calibrations on
page 2-57 for instructions.

3.

Press

Home

to return to the Ready screen. If you do not press Home, the

system automatically returns to the Ready screen after 45 seconds.

If the Auto Repeat function is turned on or Flexible Time calibration intervals are
selected and a parameter shows excessive drift, the system prints a report and
repeats the appropriate calibration for the parameter. Refer to Selecting Calibration
Frequency and Automatic Repeat
in Section 5 for more information about these
functions.

If a sensor does not reach endpoint within 90 seconds, the parameter value appears
on the screen with an asterisk and the message D5 No Endpoint also appears. The
sensor is considered not in calibration until the system completes a successful
calibration.

The metabolite one-point calibration is available only from the Analyze mode.

1

Menu Code

Procedural

Notes