D75 lamp failure – Bayer HealthCare Rapidlab 800 User Manual

Page 331

Advertising
background image

4Ć47

Troubleshooting the System

D75 Lamp Failure

Problem

The CO-ox module detects lamp levels that are inadequate for sample analysis.

Solutions

Perform the Lamp On/Off Test as described in CO-ox Optics Test, page 4-75.

Remove the lamp cover from the CO-ox module and visually verify that the
lamp is on. If the lamp is not on, replace it as described in Replacing the CO-ox
Lamp
in Section 3.

Visually verify that light is passing through the CO-ox sample chamber.

If the D code reappears, contact your Service Representative.

D75