Recalling calibration data – Bayer HealthCare Rapidlab 800 User Manual

Page 159

Advertising
background image

2Ć61

Operating the System

Recalling Calibration Data

You can print calibration data and send the data to an LIS or a data management
system. You cannot edit calibration data.

1.

Access the Recall Calibration Data screen from the Menu screen:

a. Select

4 Data Recall

and press

Enter

.

b. Select

4 Calibration Data

and press

Enter

.

2.

Select a calibration type and press

Enter

.

3.

Type the search criteria and press

Enter

.

NOTE:ĄĂ

The Calibration Date fields contain the current day’s date. If you do

not complete any other fields, the system recalls all one-point calibrations for
this day.

If you want to search . . .

Then . . .

for a selected date

type the date in both Calibration Date fields.

from the earliest date to a
specific date

leave the Calibration Date From field blank and type
the specific date in the Calibration Date To field.

from a specific date to the
present date

type the start date in the Calibration Date From field
and leave the Calibration Date To field blank.

by shift or any other
specific time period

type the start time in the Calibration Time From field
and type the end time in the Calibration Time To
field.

4.

Press

Done

.

If . . .

Then . . .

more than one calibration
is found

the Calibration Data Search Log screen appears as
shown in Figure 2-34. Continue with step 5.

one calibration is found

the Calibration Data Search Result screen appears, as
shown in Figure 2-35. Continue with step 6.

4

Menu Code

4

Advertising