Analyzing Expired Gas Samples

Use this procedure to analyze expired gas samples. During this procedure, you
inject the gas into the system. The minimum sample volume requirement is 10 mL.

BIOHAZARD:

Refer to Appendix A , Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Prepare the sample and insert the syringe in the sample port, as shown in
Figure 2-13.

Figure 2Ć13.ā Inserting an Expired Gas Sample

Sample Port

Expired Gas Sample

2.

Press

Change Sample Type

.

The Sample Type menu appears, as shown in Figure 2-14.

Figure 2Ć14.ā Sample Type Menu

Select Expired Gas.

3.

Select

Expired Gas

and press

Enter

.

4.

Press

Analyze

.

Wait for the sample pump to start.

5.

When the pump starts, slowly inject gas until the sample pump stops. Inject at
least 10 mL of gas.