800 Series Operator's Manual

NOTE:ĄĂ

If the temperature is outside of the range, verify that the power has

been on for at least 30 minutes.

40. Access the Barometer

screen from the Menu screen:

a. Select

1 Calibration

and press

Enter

.

b. Select

8 Barometer

and press

Enter

.

The Barometer Calibration screen appears.

41. Compare the displayed atmospheric pressure to the laboratory’s barometer

reading.

If . . .

Then . . .

the displayed atmospheric
pressure is correct

press

Done

.

The Ready screen appears.

the displayed atmospheric
pressure is incorrect

a. Type the correct atmospheric pressure, and press

Enter

.

b. Press

Done

to save the new atmospheric pressure

and return to the Ready screen.

c. Perform a gas two-point calibration.

42. Condition the sensors.

Refer to Conditioning the Sensors in Section 3.

CAUTION:

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

43. Install the glucose and lactate biosensors:

a. Remove

test/blank sensors (TB4).

b. Remove the lactate and glucose biosensors from the foil package.

c. Align the contacts on the biosensors with the contacts in the measurement

module.

d. Snap the body of the biosensors down into place. The contacts must be

flush with the sensors.

e. Press the tab on the spring-loaded latch down to release the latch.

44. Verify that the sensors are installed in the following order:

pO

2

pCO

2

GND

pH

Na

+

K

+

Ca

++

Cl

-

Ref

Glu

Lac

45. Perform three two-point calibrations to start hydrating the glucose and lactate

biosensors as soon as possible.

46. Allow the glucose and lactate biosensors to equilibrate for at least 30 minutes.

47. Verify that the system temperature is within the acceptable range.