Bayer HealthCare Rapidlab 800 User Manual

Page 107

Advertising
background image

2Ć9

Operating the System

CAUTION:ĄĂ

To ensure the accuracy of the CO-ox measurement, close the

CO-ox cover before pressing Analyze.

2.

Press

Analyze

.

If . . .

Then . . .

there is sufficient
sample volume

the system automatically aspirates the sample and moves it
to the measurement module.

the system prompts
you to move the
sample

a. Look to see if the sample has reached point A. If not,

turn the sample pump clockwise to move the leading
edge of the sample to point A, as shown in Figure 2-4.

b. Press

Sample in Place

.

Figure 2Ć4.ā Sample at Point A on an 850

Point A

Sample Pump

3.

When prompted, remove the sample device.

4.

Type the required information in the Patient Information screen and then press

Done

.

5.

Perform the appropriate task.

If . . .

Then . . .

the system is moving
the sample

the results appear on the screen and are updated until
analysis is complete. Continue with step 8.

you are moving the
sample

continue with step 6.

!