Sample sequence, 1ć59 learning about the system, Continued) – Bayer HealthCare Rapidlab 800 User Manual

Page 81

Advertising
background image

1Ć59

Learning About the System

Sample Sequence

Cal
Slope

Ambient Air

1 1a

2

4

3

R

S

W

7.3

6.8

Wash Clean Cal G/L

5

CO

H

SC

1.

The sample door closes and determines the sample type
and the probe extends accordingly. The waste and sample
pumps start, and the sample moves to fluid detector (FD) 1
and 1A. The CO-ox module pump moves part of the
sample to FD5.

Cal
Slope

Ambient Air

1 1a

2

4

3

R

S

W

7.3

6.8

Wash Clean Cal G/L

5

CO

H

SC

2.

The probe retracts. The sample door opens and the system
prompts you to remove the sample device. When the
sample device is removed, the door closes over the sample
port.