Rapidlab 800 Operator's Manual

Cal
Slope

Ambient Air

1 1a

2

4

3

R

S

W

• • • • • • • • • • • • • • • • • • • • • • • •

••••••••••••••

••

• • • • •

7.3

6.8

Wash Clean

5

CO

H

SC

6.

The reagent and sample pumps remain on. The
pre-segment moves through the sample port into the waste
and removes any residual reagent in the manifold.

Cal
Slope

Ambient Air

1 1a

2

4

3

R

S

W

• • • • • • • • • • • • • • • • • • • • • • •

• • • • • • • • • • •

• • • • • • •

••••••••••••••••••••

• • •

•••

••••••••••••••••••

• • •

• • • • •

7.3

6.8

Wash Clean

5

CO

H

SC

7.

The main segment moves toward the sample port. When
the main segment reaches the sample port, the sample
pump starts and moves the segment to the measurement
module until it is detected by fluid detector 1 and 1A.The
CO-ox module pump moves the 7.3 reagent to FD5. The
waste pump starts.

Cal
Slope

Ambient Air

1 1a

2

4

3

R

S

W

7.3

6.8

Wash Clean

5

CO

H

SC

8.

The probe extends and the main segment moves until the
trailing edge is at fluid detector 1. The pumps stop and the
analytes are measured. The CO-ox zero calibration is
performed.

(Continued)