D2 excessive drift – Bayer HealthCare Rapidlab 800 User Manual

Page 300

Advertising
background image

4Ć16

800 Series Operator's Manual

D2 Excessive Drift:

tHb

Problem

The tHb slope is beyond predefined limits during the calibration.

Qualifiers:

tHb

Solutions

Ensure the target value entered for the tHb slope is correct for the slope reagent
used. If required, re-enter the value and repeat the tHb slope calibration.

Deproteinize the sample path as described in Deproteinizing the Sample Path in
Section 3. Repeat the tHb slope calibration.

Check that the sample chamber is tightly in place. Refer to Cleaning the Sample
Chamber
, in Section 3. The tab on the cams should be vertical. Repeat the tHb
slope calibration.

Check that the gasket in the sample chamber is seated correctly.

If the D code reappears, contact your Service Representative.

D2