Twice weekly maintenance, Analyzing high g/l – Bayer HealthCare Rapidlab 800 User Manual

Page 180

Advertising
background image

3Ć12

Rapidlab 800 Operator's Manual

Twice Weekly Maintenance

The twice weekly maintenance schedule is based on analyzing 30 samples per day,
unless otherwise noted. If your laboratory analyzes more than 30 samples per day,
perform twice weekly maintenance more frequently.

Analyzing High G/L

Use this procedure to verify the performance of the glucose and lactate biosensors.

Materials required:

S High G/L ampule
S aspiration adapter
1.

Perform a successful two-point calibration.

2.

Prepare the High G/L ampule and insert an aspiration adapter into the sample
port and into the ampule.

3.

Press

Analyze

.

4.

When prompted, remove the sample.

5.

Type the required information in the Patient Information screen and press

Done

.

6.

Review the results.

If the glucose or lactate results are below the values recommended on the High
G/L package insert, replace the affected biosensor as described in Replacing
the Glucose and Lactate Biosensors
, page 3-86.